头部背景图片

Daily maintenance tips for high and low temperature test chambers and alternating high and low temperature test chambers

2025-01-16

Daily maintenance tips for high and low temperature test chambers and alternating high and low temperature test chambers

1. High and low temperature test chambers are generally relatively high, and we recommend placing them in a relatively benign temperature environment. Our experience temperature value is 8 ℃~23 ℃. For laboratories that do not have this condition, appropriate air conditioners or cooling towers must be equipped.

2. It is necessary to adhere to professional management by dedicated personnel. Units with conditions should periodically send dedicated personnel to the supplier's factory for training and learning, in order to gain more professional experience and ability in maintenance and repair Hongzhan Instrument.

3. Regularly clean the condenser every 3 months: For compressors that use air-cooled cooling, the condenser fan should be regularly inspected and the condenser should be cleaned and dusted to ensure good ventilation and heat transfer performance; For compressors that use water-cooled cooling, in addition to ensuring their inlet water pressure and temperature, it is also necessary to ensure the corresponding flow rate. Regular cleaning and descaling of the condenser interior is also necessary to obtain its continuous heat transfer performance.

4. Regularly clean the evaporator: Due to the different cleanliness levels of the test samples, a lot of small particles such as dust will accumulate on the evaporator under forced air circulation, and should be cleaned regularly.

5. Cleaning and balancing of circulating air blades and condenser fans: Similar to cleaning evaporators, due to the different working environments of the test chamber, many small particles such as dust may accumulate on the circulating air blades and condenser fans, and should be cleaned regularly.

6. Cleaning of waterway and humidifier: If the waterway is not smooth and the humidifier scales, it is easy for the humidifier to dry and burn, which may damage the humidifier. Therefore, it is necessary to regularly clean the waterway and humidifier.

7. After each experiment, set the temperature near the ambient temperature, work for about 30 minutes, then cut off the power and clean the inner wall of the workshop.

If the equipment needs to be relocated, it is best to do so under the guidance of technical personnel from Hongzhan Company to avoid unnecessary damage or damage to the equipment.

When the product is not in use for a long period of time, it should be powered on regularly every half month, and the power on time should not be less than 1 hour.

10. Maintenance principle:

Due to the fact that high and low temperature test chambers are mainly composed of electrical, refrigeration, and mechanical systems, once there is a problem with the equipment, a comprehensive inspection and analysis of the entire equipment system should be carried out.

Generally speaking, the process of analysis and judgment can start with "external" and then "internal", that is, after excluding external factors, the equipment can be systematically decomposed based on the fault phenomenon. Then, the system can be comprehensively analyzed and judged. Alternatively, the reverse reasoning method can be used to find the cause of the fault: first, check whether there is a problem with the electrical system according to the electrical wiring diagram, and finally check whether there is a problem with the refrigeration system. Before understanding the cause of the fault, it is not advisable to disassemble or replace components blindly to avoid unnecessary trouble.

EC-85EXT, Superior constant temperature bath (800L) Specifications

2025-01-16

EC-85EXT, Superior constant temperature bath (800L)Specifications

Project

Type

Series

EXT

Function

Temperature occurs in a way

Dry ball method

Temperature range

-70 ~ + 150 ℃

Temperature variation amplitude

Below the + 100℃

± 0.3 ℃

Above the + 101℃

± 0.5 ℃

Temperature distribution

Below the + 100℃

± 0. 7 ℃

Above the + 101℃

± 1.0 ℃

The temperature drops the time

+125 ~-55 ℃

Within 36 minutes (10℃ / minute mean temperature change)

Temperature rise time

-55 ~+125 ℃

Within 36 minutes (10℃ / minute mean temperature change)

The internal volume of the uterus was tested

800L

Test room inch method (width, depth and height)

1000mm × 800mm × 1000mm

Product inch method (width, depth and height)

1470mm × 2240mm × 2000mm

Make the material 

External outfit 

Test room control panel

machine room

Cold interductile steel plate is dark gray

Inside

Stainless steel plate (SUS304,2B polished)

Broken heat material

Test room

Hard synthetic resin

glass wool

door

Hard synthetic resin foam cotton, glass cotton

Project

Type

Series

EXT

Cooling dehumidifying device

 Cooling-down method

Mechanical section shrinkage and freezing mode and binary freezing mode

 Cooling medium

Single segment side

R404A

Binary high temperature / low temperature side

R404A / R23

Cooling and dehumidifier

Multi-channel mixed heat sink type

The condenser

(water-cooled)

Calorifier

Form

Nickel-chromium heat-resistant alloy heater

Blower

Form

Stir fan

 Controllerl

The temperature is set

-72.0 ~ +152.0 ℃

Time setting Fanny

0 ~ 999 Time 59 minutes (programmed type)
0 ~ 20000 Time 59 minutes (value)

Set decomposition energy

  Temperature  0.1 ℃ ,Time of 1 minute

Indicate accuracy

Temperature ± 0.8℃ (typ.), time ± 100 PPM

Vacation type

Value or program

Stage number

20-stage / 1 program

The number of procedures

The maximum number of incoming force (RAM) programs is 32 programs

The maximum number of internal ROM programs is 13 programs

Round-trip number

Maximum of 98 or unlimited

Number of round-trip repeats

Maximum 3 time

Displace the end

Pt 100Ω ( at 0 ℃ ),grade ( JIS C 1604-1997 )

Control action

When splitting the PID action

Endovirus function

Early delivery function, standby function, setting value maintenance function, power outage protection function,

Power action selection function, maintenance function, transportation round-trip function,

Time delivery function, time signal output function, overrising and overcooling prevention function,

Abnormal representation function, external alarm output function, setting paradigm representation function,

Transport type selection function, the calculation time represents the function, the slot lamp lamp function

Project

Type

Series

EXT

Control panel

Equipment machine

LCD operating panel (type contact panel),

Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,

Time signal output terminal, power cord connector

 Protective device

Refrigerating cycle

Overload protection device, high blocking device

Calorifier

Temperature over-rise protection device, temperature fuse

Blower

Overload protection device

Control panel

Leakage breaker for power supply, fuse (heater,),

Fuse (for operating loop), temperature rise protection device (for testing),

Temperature rise overcooling prevention device (test material, in microcomputer)

Pay belongs to the product

Test material shed shed by * 8

Stainless steel shed (2), shed receiving (4)

Fuse

Operating loop Protection Fuses (2)

Operating specification

(1 )

 Else

Bolus (Cable hole: 1)

Equipment products

Adventitia

Hard borosilicate glass270mm× 190mm

 

Cable hole

Bore size 50mm

 

The trough inside the lamp

AC100V 15W White hot ball

 

Wheel

 

6

 

Horizontal adjustment

 

6

 

Electrovirus characteristics

Source  * 5.1

AC Three-phase  380V  50Hz

Maximum load current

60 A

Capacity of the leakage breaker for the power supply

80A

Sensory current  30mA

Power distribution thickness

60mm2

Rubber insulation hose

Coarseness of grounding wire

14mm2

 Cooling water * 5.3

Water yield

5000 L /h (When the cooling water inlet temperature is 32℃)

Water pressure

0.1 ~ 0.5MPa

Side pipe diameter of the device

PT1 1/4

 Tubing

 drain-pipe   * 5.4

PT1/2

Product weight

700kg

 

EC-85MHPM-W, High load corresponding constant temperature and humidity tank (800L)

2025-01-16

EC-85MHPM-W, High load corresponding constant temperature and humidity tank (800L)

Project

Type

Series

MHPM-W

Function

Temperature and humidity mode

The way of wet ball

Temperature range

-40 ~ + 100 ℃

Humidity range

20 ~ 98%RH

(According to the anaphase 3 items)

Changes in temperature and humidity

± 0.3 ℃ / ±2.5%RH

Temperature and humidity distribution

± 0.5 ℃ / ±5.0%RH

The temperature drops the time

+20 ~ -40 ℃

Within 75 minutes

Temperature rise time

-40 ~ + 100 ℃

Within 50 minutes

The internal volume of the uterus was tested

800 L

Test room inch method (width, depth and height)

1000mm × 800mm × 1000mm

Product inch method (width, depth and height)

1400mm × 1190mm × 1795mm

Make the material

External outfit

Test room control panel

machine room

Cold steel plate, cold steel plate beige

(Color table 2.5Y8 / 2)

Inside

Stainless steel plate (SUS304,2B polished)

Broken heat material

Test room

Hard synthetic resin

Door

Hard synthetic resin foam cotton, glass cotton

Project

Type

Series

MHPM-W

Cooling removal, wet device

 Cooling-down method

Mechanical section shrinkage mode

 Cooling medium

R404A

The oneself can shrink machine

Output (number of staff)

1.5kW (1)

Cooling and dehumidifier

Multi-channel mixed heat sink type

The condenser

Multi-channel mixed radiator sink (air cooling)

Calorifier

Form

Nickel-chromium heat-resistant alloy heater

Volume

3.5kW

  Humidifier

Form

Steam generation

Volume

1.8Kw×2

Blower

Form

Multi-channel mixed radiator sink (air cooling)

Motor capacity

40W

Feed water unit

The water supply cylinder Water supply method

Water quality

Pure water * Automatic water supply

("Please refer to the automatic water supply.")

Volume

 Gravity type

Moisturizing disk

 Gravity type

Controller

Temperature setting range

-42.0 ~ + 102.0 ℃

Humidity setting range

0 ~ 98%RH (Dry Bulb Temperature 10 ~ 85 ℃ )

Time set range

0 ~ 999Time of 59 min(Program setting type) 0 ~ 20000 Time of 59 min(The value type)

Set decomposition energy

Temperature 0.1℃, humidity 1% RH for 1 min

Indicate accuracy

Temperature ± 0.8℃ (tp.), humidity ± 1% RH (tp.), time ± 100 PPM

Vacation type

Value or program

Stage number

20-stage / 1 program

The number of procedures

The maximum number of incoming force (RAM) programs is 32 programs

The maximum number of internal ROM programs is 13 programs

Round-trip number

 98 times maximum or unlimited

Number of round-trip repeats

Maximum 3 heavy

Displace the end

Pt 100Ω ( at 0 ℃ ), grade B( JIS C 1604-1997 )

Control action

When splitting the PID action

Internal function

Early delivery function, standby function, setting value maintenance function, power outage protection function,

Power action selection function, maintenance function, transportation round-trip function,

Time delivery function, time signal output function, overrising and overcooling prevention function,

Abnormal representation function, external alarm output function, setting paradigm representation function,

Transport type selection function, the calculation time represents the function, the slot lamp lamp function

Project

Type

Series

MHPM-W

Control panel

Equipment machine

LCD operating panel (type contact panel),

Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,

Time signal output terminal, power cord connector

 Protective device

Refrigerating cycle

Overload protection device, high blocking device

Calorifier

Temperature over-rise protection device, temperature fuse

Humidifier

Air burning prevention device, humidifying disc water level regulator

Blower

Overload protection device

Control panel

Leakage breaker for power supply, fuse (for heater, humidifier),

Fuse (for operating loop), temperature rise protection device (for testing),

Temperature rise overcooling prevention device (test material, in microcomputer)

Offproducts (sets)

House receiver (4), house board (2), wet ball wick (15), operation manual (1)

Equipment products

Adventitia

Hard borosilicate glass  800mm× 800mm

2

Cable hole

Bore size 50mm

The trough inside the lamp

AC100V 15W White hot ball

Wheel

 

Horizontal adjustment

 

Electrovirus characteristics

Source 

AC   three-phase 380V 50Hz

Maximum load current

25A

Capacity of the leakage breaker for the power supply

50A

Sensory current 30mA

Power distribution thickness

14mm2

Rubber insulation hose

Coarseness of grounding wire

5.5mm2

Tubing

Drain-pipe

PT1/2

Product quality

550kg

ER-135MHP-W, Step in constant temperature and humidity bath, constant temperature and humidity chamber style book (water-cooled)

2025-01-16

ER-135MHP-W, Step in constant temperature and humidity bath, constant temperature and humidity chamber style book (water-cooled)

Project

Type

Form

ER - 135MHP-W

 

Nature

ability

*1

*2

Temperature and humidity mode

The way of wet ball

Temperature and humidity model * 3

-40~+ 80 ℃ /10~95%RH

Temperature and humidity electrode amplitude

± 0.3 ℃ / ± 2.5%RH

Temperature and humidity distribution

± 0.75 ℃ / ± 5.0%RH

The temperature drops the time

+20~ -40 ℃ Within 170 minutes

Temperature rise time

+20~+ 80 ℃ Within 50 minutes

Noise value

*6

Room outside

65dB

Indoor side

79B

Statutory freezing capacity ( 50 / 60Hz )

2.57/2.81

Product shape-inch method (width, depth and height)

Please refer to the configuration diagram

 

 

 

Warehouse board laboratory

Bed area

13 . 0 m2

 

The shape of inch method

*7

Wide

3600mm

Deep

3600mm

High

2325mm

Have oneself

Intensification method

*7

Wide

3450mm

Deep

3450mm

High

2100mm

External materials

Ivory painted steel plate

Inside the material

Stainless steel plate (SUS304)

Broken heat material

Hard synthetic resin

The weight resistance of the board

5.9 k N/ m2 {600 kgf / m2 }

Door (width and height)

830 mm × 1800 mm Single piece open

Measurement (width and height)

190 mm × 320 mm The door

Indoor light capacity (quantity)

60W( 2 individual )

Cable hole

φ 50 1 (500mm from bed surface height, 1 side)

Controller model and shape method

Wide, deep and high

EU - 6 5MH*2 unit 1200 mm × 710 mm × 1940 mm

 

External outfit

 Indoor

stainless steel sheet and plate (SUS304)

 Outdoor

L coating color on steel plate; ivory color: color table mark 2.5Y8/2 )

Operate

Steel plate coating color; ivory color: color table mark2.5Y8/2

 

The oneself can shrink machine

Cooling medium

R404A

Output of freezer 1 (staff)

2.2kW(1) × 2

Cooling and dehumidifier

Multi-channel mixed heat sink type

Condenser (staff number)

Transverse fin type (2) 2

Calorifier

Form

Nickel-chromium heat-resistant alloy heater

Volume

5 . 7kW(3 φ 200V) × 2

Humidifier

Form

SUS 316L Make a surface steam type humidifier

Volume

5 . 4kW(3 φ 200V) × 2

Air supply device

Form (quantity)

Single air supply multi-blade air fan (1)

Survival outside the machine

0 Pa{ 0 mmAq}

Blowing rate

25 (m3 / min )

Back to the uterus

1100( r / min )

Motor form

VTFO-K

Motor output (pole number)

0.75kW(6) × 2

Moisturizing water supply

installation 

The water supply cylinder

Water quality * 8

Electrical conductivity is below 10 μ S / cm

Volume

4L × 2

Feedwater

way

Moisturizing disk

 gravity type

Wet ball Wick disc

Electromagnetic valve type

Controller

The temperature is set

-42.0 ~+ 82.0 ℃

Humidity is set

0~98% RH ( Dry Bulb Temperature 10 ℃ ~ 80 ℃ )

Time setting Fanny

0 ~ 999 Time 59 points (programmed type)0 ~ 20000 Time 59 points (value)

Set decomposition energy

Temperature 0.1 ℃ , humidity 1% RH , time 1 minute

Indicate accuracy

 Temperature  ± 0.8 ℃ ( typ. ), humidity ± 1% RH ( typ. ),  time ± 100 PPM

Vacation type

Value or program

Stage number

20 stage / 1 form

The number of procedures

Number of incoming force (RAM) programs, maximum 32 programs / internal ROM programs, maximum 13 programs

Round-trip number

Maximum of 98 or unlimited

Number of round-trip repeats

Maximum 3 time

Displace the end

JP t 100Ω (a t 0 ℃ ), grade B(JIS C1604-1997 )

control action

When splitting the PID action

Endovirus function

Early delivery function, standby function, setting value maintenance function, power outage protection function,

Power action selection function, maintenance function, transportation round-trip function,

Time delivery function, time signal output function, overrising and overcooling prevention function,

Abnormal representation function, external alarm output function, setting paradigm representation function,

Transport type selection function, the calculation time represents the function, the slot lamp lamp function

Project

Type

Form

ER - 55MHP-W

Temperature control machine

Control panel

Equipment machine

Color LCD operating panel (force contact mode),

Represents the lamp (power supply, transport, abnormal), test material power supply system royal terminal,

External alarm terminal, time signal output terminal, power cord connector,

 Protective device

Warehouse plate test uterine cavity

Test material with

Temperature rise overcooling prevention function (microcomputer: automatic setting)

Temperature rise ter (set: Set above 60℃)

Refrigerating cycle

Overload protection device, high blocking device

Calorifier

Temperature over-rise protection device, temperature fuse

  Humidifier

Air burning prevention device (2 weight), wet plate water level regulator

Blower

Overload protection device

Control panel

Leakage breaker for power supply, fuse (for heater, humidifier),

Fuse (for operating loop), fan reverse prevention relay

In addition to control

Frost system

Defrosting method * 5

Stop cycling defrosting (freezer transport stop),

Heat up defrosting

Cold control

But the system

water

 

Concondensation control mode (quantity)

 

Water-making valve for cooling water (2) 2

Repeat

measure

Therstat (operated)

300kg × 2

Electricity

Endyceps sinensis

particular

nature

Source

Communication  three phases 380V 50Hz

Maximum load current

45A × 2

The power supply uses a leakage breaker

60A ( Sensory current 30mA) × 2

Power distribution

Rubber insulation hose 22m m2 × 2

Coarseness of grounding wire

5.5m m2 × 2

cold

step back

water

*9

 Water yield;the yield of water

40L / min (  inlet water temperature  32 ℃ ) × 2

Water pressure

0.1~0.5M Pa{ 1.0~5.0 kgf / cm2 }

 water temperature

18~ 32 ℃

Pipe-inch method

Inlet / exit

PT 1 parent seat / PT 1 parent seat

Humidifying water supply pipe distribution inch method * 8

PT 1 / 2 parent seat

arrange

water

Pipe-inch method

drain-pipe

PT 1 / 2 parent seat

Its

 

Product products (staff)

 reduce pressure(1) * 8

Wet ball wick (15)

Take instructions (1)

High, low temperature, and low pressure test chamber

2025-01-16

High, low temperature, and low pressure test chamber

The low-pressure test chamber is mainly used in the fields of aviation, aerospace, information, electronics, etc., to determine the environmental adaptability and reliability tests of instruments, electrical products, materials, components, and equipment under single or simultaneous effects of low-pressure, high-temperature, and low-temperature, and to measure the electrical performance parameters of the test piece when energized.

Product principle and function

The high, low temperature, and low pressure test chamber is mainly used in aviation, aerospace, electronics, national defense, scientific research, and other industrial sectors to determine the storage and transportation reliability test of electrical and electronic products (including components, materials, and instruments) under the single or simultaneous action of high, low temperature, and low pressure. It can also be used to test the electrical performance parameters of the specimens when they are energized.

Editing of structural device for high, low temperature and low pressure test chamber

1. The outer shell is made of cold-rolled steel plate sprayed with plastic or stainless steel, and the inner shell is made of stainless steel

2. Thermal insulation layer: ultra-fine glass wool and hard polyurethane

3. Cooling method: Compression cooling method (air-cooled condenser)

4. Refrigerator: Original French Taikang fully enclosed compressor unit

5. Heater: grade stainless steel fin heater

6. Convection system: Multi blade fan, air conditioning dedicated motor

7. Temperature controller: Digital intelligent temperature controller, using PT-l00 sensor for temperature control.

8. The arrangement of electrical components in the control cabinet is reasonable, the wiring is neat, and labeled sheathed wiring terminals are used.

9. Castor solenoid valve, Danfoss expansion valve

10. High quality vacuum pump

Editing of vacuum system for high, low temperature and low pressure test chamber

1. Composition of vacuum system: The vacuum system is a component that obtains and measures the vacuum degree, consisting of a vacuum measurement system and a vacuum acquisition unit

2. Vacuum measurement: Vacuum measurement is carried out using a pressure sensor. This vacuum gauge uses pressure strain to achieve vacuum measurement, and the measurement data is a linear electrical signal that can be directly input into the device controller for display and control

environment condition

Temperature: 5 ℃~+28 ℃ (average temperature within 24 hours ≤ 28 ℃)

Relative humidity: ≤ 85% RH

Air pressure: 86kPa~106kPa

Power supply conditions: three-phase four wire+protective ground wire, voltage range: AC (380 ± 38) V

Allowable frequency fluctuation range: (50 ± 0.5) Hz

The grounding resistance of the protective grounding wire is less than 4 Ω

Technical specifications of high, low temperature and low pressure test chamber

1. Studio size:

FA-4 4 cu ft (113L)

-73°C to +177°C

N/A  

FA-10  10 cu ft (283L)

-73°C to +177°C

20% to 95%RH  

FA-16  16 cu ft (453L)

-73°C to +177°C

20% to 95%RH  

FA-35  35 cu ft (991L)

-73°C to +177°C

20% to 95%RH  

FA-64  64 cu ft (1812L)

-73°C to +177°C

20% to 95%RH  

FA-96  96 cu ft (2718L)

-73°C to +177°C

20% to 95%RH  

2. Temperature range: -40~150 ℃

3. Temperature deviation: ± 2 ℃

4. Temperature fluctuation: ± 0.5 ℃

5. Temperature uniformity: ≤ 2 ℃

6. Air pressure level: 4-84kpa

7. Heating rate 1.0-3.0 ℃/min

8. Cooling rate 0.7-1.0 ℃/min

9. Control system controller Siemens PLC controller, Siemens color LCD display touch screen

10. Accuracy Range Setting Accuracy: Temperature ± 0.1 ℃, Indication Accuracy: Temperature ± 0.1 ℃, Resolution: ± 0.1 ℃

11. Temperature sensor platinum resistance PT100 Ω/MV

12. Pressure Sensor Electronic Pressure Transmitter

13. Fully independent heating system, nickel chromium alloy electric heating heater

14. Refrigeration system: original French "Taikang" fully enclosed air-cooled single-stage/cascade compressor refrigeration method

15. Temperature resistant and low-noise air conditioning motor for circulation system. Multi blade centrifugal fan

Safety protection device

1. Overload and short circuit protection

2. Overtemperature protection

3. High and low pressure protection of refrigeration unit

4. Alarm sound prompt

Description of Controller Q8-902 for High Low Temperature and Low Pressure Test Chamber

Operating interface

Touchscreen Chinese interface, with digital display, real-time curve, operation selection, timing setting, program setting, alarm, parameter setting and other interfaces

Curve recording function

Can save the set values, sampling values, and sampling time of the device; The maximum storage time is 30 days. Simultaneously storing 5 real-time curves with different capacities and resolutions for 30 days, 7 days, 24 hours, 8 hours, and 1 hour, and displaying real-time data to meet different user needs

Printing function

Connectable micro printer (optional)

Program capacity

Program: Up to 30 curves

Program segment: Each program can have up to 50 segments (total program segment 1200 segments)

Program segment grouping: Each curve can be further grouped into segments according to any continuous segment, up to a maximum of 10 segments can be combined, and segments are allowed to overlap

Internal loop: The program segments within the program are looped in sections as the minimum loop unit, with a maximum of 99 loops per section

Program loop: can loop up to 99 times

Program connection: After the program runs, it can connect to the selected next program

Related standards

In addition to the content specified in this technical requirement, the equipment meets the requirements of GB/T10591 and GB/T10592.

Data and Services

1. Provide technical information such as electrical schematics, consumable parts lists, operation and maintenance manuals for the equipment, as well as technical information for the main purchased supporting parts.

2. Train 1-2 equipment operation and maintenance personnel to fully master the operation skills and general repair and maintenance skills of the equipment.

Within one year from the date of acceptance of the equipment, free after-sales service shall be provided to the buyer, and after one year, accessories of the equipment shall be provided to the buyer at cost price.

Industrial Computer Reliability Testing

2025-01-16

Industrial Computer Reliability Testing

Industrial computers can be divided into three categories according to their application attributes:

(1) Board class: includes Single Board Computer (SBC), Embedded board (Embedded Board), Black Plane, PC/104 module. (2) Subsystem class: includes single-board computers, boards, chassis, power supplies and other peripherals combined into operational subsystems, such as industrial servers and workstations. (3) System integration solutions: refers to a set of systems developed for a professional field, including the required software and hardware and surrounding, such as automatic teller machines (ATMs). The application of industrial computers widely covers ATM, POS, medical electronic equipment, game machines, gambling equipment, etc. The multi-field industry makes industrial computers must be able to withstand the use of sunlight, high and low temperature, wet and other environments, so the relevant reliability test is the focus of various manufacturers in the research and development test.

Common reliability tests for industrial computers:

(1) Wide temperature test

According to the actual application environment can be divided into four categories: 1. Outdoor: especially for the extreme low temperature or high temperature areas, such as northern Europe and desert countries, the temperature range can be from -50 to 70°C; 2. Enclosed space: for example, where heat sources are generated, such as next to a boiler, the high temperature range is about 70°C; 3. Mobile equipment: such as vehicle equipment, the high temperature can be up to 90°C according to the car area; 4. Special harsh environment: such as aerospace equipment, military, oil drilling equipment.

High and Low Temperature Test Chamber

(2) Aging stress test

The temperature range is from -40°C to 85°C, and the temperature variation rate is 10 °C per minute for cyclic testing

(3) No wind high temperature test

At present, in order to prevent dust, industrial computers are planned to be closed and fanless in the mechanism design, so more and more manufacturers begin to pay attention to high temperature testing in a windless environment to ensure that high temperatures will not collapse

Temperature and Humidity Test Chamber

Note: For complete industrial computer test conditions, please consult LAB COMPANION

 

Inverter- Reliability Test

2025-01-16

Inverter- Reliability Test

Inverter- reliability test also known as voltage converter, its function is to convert DC low voltage into AC high voltage, some electronic equipment must be driven by AC power, but we provide is DC power, at this time you must use the Inverter, direct current into alternating current to drive the electronic parts. Inverter- reliability test also known as voltage converter, its function is to convert DC low voltage into AC high voltage, some electronic equipment must be driven by AC power, but we provide is DC power, at this time you must use the Inverter, direct current into alternating current to drive the electronic parts.

Relevant test conditions:

Item

temperature

time

other

Initial test at normal temperature

25 ℃

TIME≥2 hours

-

Low temperature initial test

0 ℃ or -5 °C

TIME≥2 hours

-

High temperature initial test

60℃

TIME≥2 hours

-

High temperature and high humidity test

40℃/95%RH

240 hours

-

High temperature storage test

70℃

TIME≥96 hours or 240 hours

-

Low temperature storage test -1

-20°C

TIME≥96 hours

-

Low temperature storage test -2

-40℃

240 hours

-

High temperature and high humidity storage test

40℃/90%RH

TIME≥96 hours

-

Temperature cycle test

-20℃~ 70℃

5 cycle

Room temperature ↓-20 ℃(4 hours)↓ Room temperature (90%RH.4 hours)↓70°C(4 hours)↓ Room temperature (4 hours)

High temperature load test

55 ℃

equivalent load, 1,000 hours

-

Life test

40°C

MTBF≥40000 hours

-

on/off test (power cycle)

-

-

1min:on, 1min:off, 5,000 cycles using equivalent load

Vibration test

-

-

Acceleration 3q, frequency 10~55HZ, X, Y, Z three directions 10 minutes each, a total of 30 minutes

Impact test

-

-

Acceleration of 80g, 10 ms each time, Three times in X, Y, Z directions

Note 1: The tested module should be placed at normal (15~35° C,45~65%RH) for one hour before testing

Applicable equipment:

1. High and low temperature test chamber

High and Low Temperature Test Chamber

2. High temperature and high humidity test chamber

High Temperature and High Humidity Test Chamber

3. Rapid temperature cycle test chamber

 Rapid Temperature Cycle Test Chamber

 

 

 

 

 

 

 

 

Large volume vertical temperature shock test chamber

2025-01-16

Large volume vertical temperature shock test chamber

By alternating the movement of the test specimen between the cold chamber and the hot chamber, the temperature shock test chamber can apply a very, very fast rate of temperature change to the test specimen, which can expose possible defects and also be used to define the service life of the test specimen.

The temperature shock test chamber of Lab Companion TS-2 model is mainly used for heavy loads. Due to its compact structure and small footprint, it is particularly recommended for use in production lines.

Lab Companion has a research institution specializing in the development of environmental testing equipment, with mature environmental testing research methods and laboratories. It has gathered a group of excellent talents and well-known experts in the industry, and a strong R&D team leads the development direction of domestic environmental testing technology. Currently, the company has independent intellectual property rights for environmental testing equipment, reliability testing equipment, high and low temperature test chambers, high and low temperature humidity test chambers, constant temperature and humidity test chambers, rapid temperature change test chambers, cold and hot impact test chambers, three comprehensive test chambers, high and low temperature and low pressure test chambers, solar radiation test chambers, industrial ovens, cold and hot impact test chambers, walk-in constant temperature and humidity test chambers, environmental stress screening test chambers, walk-in constant temperature and humidity test chambers, high and low temperature impact test chambers. Box, constant temperature and humidity testing machine, constant temperature and humidity testing box, solar radiation testing box, High and low temperature humidity test chamber, temperature and humidity control chamber, UV accelerated aging test machine, UV accelerated weather resistance test machine, walk-in test chamber, walk-in environmental test chamber, walk-in high and low temperature test chamber, temperature and humidity control test chamber, UV climate resistance test chamber, UV aging tester, high and low temperature low-pressure test chamber, rapid temperature cycling test chamber, walk-in constant temperature and humidity test chamber, walk-in high and low temperature humidity test chamber, precision oven, programmable constant temperature and humidity test chamber, programmable constant temperature and humidity test machine, xenon lamp aging test chamber, high and low temperature alternating humidity test chamber, constant temperature and humidity test chamber, as well as walk-in high and low temperature humidity test chamber and high wind speed rain test chamber and other climate and environmental testing equipment and customization The product is at the forefront domestically and at the international first-class level. Welcome new and old customers to contact us for inquiries. We will be dedicated to serving you!

Polarizer Test Conditions

2025-01-16

Polarizer Test Conditions

Polarizer is one of the basic parts of the liquid crystal display, is a light plate that only allows a certain direction of light to pass through, in the process of making the liquid crystal plate, must be used above and below each piece, and into the staggered direction placed, mainly used for electric field and no electric field when the light source produces a phase difference and the state of light and dark, to display subtitles or patterns.

Relevant test conditions:

Because the molecular structure of iodine is easy to destroy under high temperature and humidity conditions, the durability of the polarizer produced by iodine dyeing technology is poor, and generally can only meet:

High temperature: 80℃×500HR

Hot and humid: working conditions below 60℃×90%RH×500HR

However, with the expansion of the use of LCD products, the wet and hot working conditions of polarizing products are becoming more and more demanding, and there has been a demand for polarizing plate products that work at 100 ° C and 90%RH conditions, and the highest conditions at present are:

High temperature: 105℃×500HR

Humidity and heat: test requirements below 90℃×95%RH×500HR

The durability test of polarizer includes four test methods: high temperature, wet heat, low temperature and cold and heat shock, of which the most important test is the wet and heat test. The high temperature test refers to the high temperature working conditions of the polarizer at a constant baking temperature. At present, according to the technical grade of the polarizer, it is divided into:

Universal type: working temperature is 70℃×500HR;

Medium durability type: the working temperature is 80℃×500HR;

High durability type: the operating temperature is 90℃×500H above these three grades.

Because the basic materials of the polarizing film PVA film and iodine and iodide are easily hydrolyzed materials, but also because the pressure sensitive adhesive used in the polarizing plate is also easy to deteriorate under high temperature and high humidity conditions, the most important things in the environmental test of the polarizing plate are high temperature and wet heat.

Temperature Cycling Test Chamber

 

 

Purpose of Temperature Shock Test

2025-01-16

Purpose of Temperature Shock Test

Reliability environmental test In addition to high Temperature, low temperature, high temperature and high humidity, temperature and humidity combined cycle, temperature Shock (cold and hot Shock) is also a common test project, temperature shock Testing (Thermal Shock Testing, Temperature Shock Testing, referred to as: TST), the purpose of the temperature shock test is to find out the design and process defects of the product through the severe temperature changes that exceed the natural environment [temperature variability greater than 20℃/min, and even up to 30 ~ 40℃/min], but there is often a situation where the temperature cycle is confused with the temperature shock. "Temperature cycle" means that in the process of high and low temperature change, the temperature change rate is specified and controlled; The temperature change rate of "temperature shock" (hot and cold shock) is not specified (Ramp Time), mainly requires Recovery Time, according to the IEC specification, there are three kinds of temperature cycle test methods [Na, Nb, NC]. Thermal shock is one of the three [Na] test items [rapid temperature change with specified conversion time; medium: air], the main parameters of temperature shock (thermal shock) are: High temperature and low temperature conditions, residence time, return time, number of cycles, in high and low temperature conditions and residence time the current new specification will be based on the surface temperature of the test product, rather than the air temperature in the test area of the test equipment.

Thermal shock test chamber:

It is used to test the material structure or composite material, in an instant under the continuous environment of extremely high temperature and extremely low temperature, the degree of tolerance, so as to test the chemical changes or physical damage caused by thermal expansion and contraction in the shortest time, the applicable objects include metal, plastic, rubber, electronic.... Such materials can be used as the basis or reference for the improvement of its products.

Thermal Shock Test Chamber

The cold and thermal shock (temperature shock) test process can identify the following product defects:

Different expansion coefficient caused by the stripping of the joint

Water enters after cracking with different expansion coefficient

Accelerated test for corrosion and short circuit caused by water infiltration

According to the international standard IEC, the following conditions are common temperature changes:

1. When the equipment is transferred from a warm indoor environment to a cold outdoor environment, or vice versa

2. When the equipment is suddenly cooled by rain or cold water

3. Installed in the outside airborne equipment (such as: automobile, 5G, outdoor monitoring system, solar energy)

4. Under certain transport [car, ship, air] and storage conditions [non-air-conditioned warehouse]

Temperature impact can be divided into two types of two-box impact and three-box impact:

Instructions: Temperature impact is common [high temperature → low temperature, low temperature → high temperature] way, this way is also called [two-box impact], another so-called [three-box impact], the process is [high temperature → normal temperature → low temperature, low temperature → normal temperature → high temperature], inserted between the high temperature and low temperature, to avoid adding a buffer between the two extreme temperatures. If you look at the specifications and test conditions, there is usually a normal temperature condition, the high and low temperature will be extremely high and very low, in the military specifications and vehicle regulations will see that there is a normal temperature impact condition.

IEC temperature shock test conditions:

High temperature: 30, 40, 55, 70, 85, 100, 125, 155℃

Low temperature: 5, -5, -10, -25, -40, -55, -65℃

Residence time: 10min, 30min, 1hr, 2hr, 3hr(if not specified, 3hr)

Temperature shock residence time description:

The Dwell Time of temperature shock in addition to the requirements of the specification, some will depend on the weight of the test product and the surface temperature of the test product

The specifications of the thermal shock residence time according to weight are:

GJB360A-96-107, MIL-202F-107, EIAJ ED4701/100, JASO-D001... Let's wait.

The thermal shock residence time is based on the surface temperature control specifications: MIL-STD-883K, MIL-STD-202H(air above the test object)

MIL883K-2016 requirements for [temperature shock] specification:

1. After the air temperature reaches the set value, the surface of the test product needs to arrive within 16 minutes (residence time is not less than 10min).

2. High temperature and low temperature impact are more than the set value, but not more than 10℃.

Follow-up action of IEC temperature shock test

Reason: The IEC temperature test method is best considered as part of a series of tests, because some failures may not be immediately apparent after the test method is completed.

Follow-up test items:

IEC60068-2-17 Tightness test

IEC60068-2-6 Sinusoidal vibration

IEC60068-2-78 Steady Humid heat

IEC60068-2-30 Hot and humid temperature cycle

Tin whisker (whisker) temperature impact test conditions finishing:

1. - 55 (+ 0 / -) 10 ℃ please - 85 (+ / - 0) 10 ℃, 20 min / 1 cycle (500 cycle check again)

1000 cycles, 1500 cycles, 2000 cycles, 3000 cycles

2. 85(±5)℃←→-40(+5/-15)℃, 20min/1cycle, 500cycles

3.-35±5℃←→125±5℃, dwell for 7min, 500±4cycles

4. - 55 (+ 0 / -) 10 ℃ please - 80 (+ / - 0) 10 ℃, 7 min reside, 20 min / 1 cycle, 1000 cycles

Thermal shock testing machine product features:

Defrosting frequency: defrosting every 600cycles [Test condition: +150℃ ~ -55℃]

Load adjustment function: The system can automatically adjust according to the load of the product to be tested, without manual setting

High weight load: Before the equipment leaves the factory, use aluminum IC(7.5Kg) for load simulation to confirm that the equipment can meet the demand

Temperature shock Sensor location: The air outlet and return air outlet in the test area can be selected or both can be installed, which conforms to the MIL-STD test specification. In addition to meeting the requirements of the specification, it is also closer to the impact effect of the test product during the test, reducing the test uncertainty and distribution uniformity.

High and Low Temperature Shock Test Chamber

Tags