头部背景图片

EC-35EXT, Superior constant temperature bath (306L)

EC-35EXT, Superior constant temperature bath (306L)

Project

Type

Series

EXT

Function

Temperature occurs in a way

Dry wet bulb method

Temperature range

-70 ~ +150 ℃

Range of temperature

Below the + 100℃

±0.3 ℃

Above the + 101℃

±0.5 ℃

Temperature distribution

Below the + 100℃

±0. 7 ℃

Above the + 101℃

±1.0 ℃

The temperature drops the time

+125 ~-55 ℃

Within 18 points (10℃ / point average temperature change)

Temperature rise time

-55 ~+125 ℃

Within 18 minutes (10℃ / minute)

The internal volume of the uterus was tested

306L

Test room inch method (width, depth and height)

630mm × 540mm × 900mm

Product inch method (width, depth and height)

1100mm × 1960mm × 1900mm

Make the material

External outfit

Test room control panel

machine room

Cold interductile steel plate is dark gray

Inside

Stainless steel plate (SUS304,2B polished)

Broken heat material

Test room

Hard synthetic resin

door

Hard synthetic resin foam cotton, glass cotton

Project

Type

Series

EXT

Cooling dehumidifying device

Cooling-down method

Mechanical section shrinkage and freezing mode and binary freezing mode

Cooling medium;coolant

Single segment side

R 404A

Binary high temperature / low temperature side

R 404A / R23

Cooling and dehumidifier

Multi-channel mixed heat sink type

The condenser

(water-cooled)

Calorifier

Form

Nickel-chromium heat-resistant alloy heater

Blower

Form

Stir fan

Controller

The temperature is set

-72.0 ~ + 152.0 ℃

Time setting Fanny

0 ~ 999 Time 59 minutes (formula) 0 ~ 20000 Time 59 minutes (formula formula)

Set decomposition energy

Temperature was 0.1℃ for 1 min

Indicate accuracy

Temperature ± 0.8℃ (typ.), time ± 100 PPM

Vacation type

Value or program

Stage number

20-stage / 1 program

The number of procedures

The maximum number of incoming force (RAM) programs is 32 programs

The maximum number of internal ROM programs is 13 programs式

Round-trip number

Max. 98, or unlimited

Number of round-trip repeats

Maximum 3 times

Displace the end

Pt 100Ω ( at 0 ℃ ),grade ( JIS C 1604-1997 )

Control action

When splitting the PID action

Endovirus function

Early delivery function, standby function, setting value maintenance function, power outage protection function,

Power action selection function, maintenance function, transportation round-trip function,

Time delivery function, time signal output function, overrising and overcooling prevention function,

Abnormal representation function, external alarm output function, setting paradigm representation function,

Transport type selection function, the calculation time represents the function, the slot lamp lamp function

Project

Type

Series

EXH

Control panel

Equipment machine

LCD operating panel (type contact panel),

Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,

Time signal output terminal, power cord connector

 Protective device

Refrigerating cycle

Overload protection device, high blocking device

Calorifier

Temperature over-rise protection device, temperature fuse

Blower

Overload protection device

Control panel

Leakage breaker for power supply, fuse (heater,),

Fuse (for operating loop), temperature rise protection device (for testing),

Temperature rise overcooling prevention device (test material, in microcomputer)

Pay belongs to the product

Test material shed shed by * 8

Stainless steel Shshed (2), shed (4)

Fuse

Operating loop Protection Fuses (2)

Operating specification

( 1 )

 Else

Bolus (Cable hole: 1)

Equipment products

Adventitia

Heat-resistant glass: 270mm: 190mm

1

 

Cable hole

Inner diameter of 50mm

1

 

The trough inside the lamp

AC100V 15W White hot ball

1

 

Wheel

 

6

 

Horizontal adjustment

 

6

 

Electrovirus characteristics

Power supply is * 5.1

 AC Three-phase  380V  50Hz

Maximum load current

60A

Capacity of the leakage breaker for the power supply

80A

Sensory current  30mA

Power distribution thickness

60mm2

Rubber insulation hose

Coarseness of grounding wire

14mm2

Cooling water at * 5.3

Water yield

5000 L /h (When the cooling water inlet temperature is 32℃)

water pressure

0.1 ~ 0.5MPa

Side pipe diameter of the device

PT1 1/4

 Tubing

Drain-pipe  * 5.4

PT1/2

Product weight

700kg

Tags